Pyrgeometer SGR4_0133

Summary



TypePyrgeometer
VendorKipp and Zonen (OTT Hydromet)
ModelKipp and Zonen SGR4 pyrgeometer
S/N210133
ID Prefixpyrgeo
CORIOLIX IDpyrgeo210133
ClassAtmospheric
LocationAbove SCR
Conditionexcellent
ConfigurationMounted in CV4 Ventilation Unit
Native Sample Rate1.0 Hz
UDP Port(s)53136 None
EnabledTrue
 


Image Not Available
 
 

General Description



The SGR4-V has a digital interface and is designed for meteorological observations. The excellent thermal stability of the dome construction eliminates the need for dome temperature measurements or dome shading.

 
 

Data Collection Specs

Native Data Feed
Signal Type Digital
Communication Type RS-485
Serial Details
Baud Rate 19200
Parity E
Stop Bits 1
Byte Size 8
Timeout (s) None
Polled Start Commands None
Polled Serial Commands
Ethernet Details
MAC Address
IP Address 10.3.0.136
Protocol udp
Source File Details
Path
Filename
Native Data Format
Native Data Description Modbus RTU Protocol. Example Data Line: radiom210133 rad_sgr4 2022-11-05T00:00:00.6009Z 611,133,-14.18,262.67,-8.81,12.05,1,1,0 ***** Field Descriptions: Coriolix Sensor ID, LDS Logger ID, LDS Time Stamp [UTC], Device_Type: [611 = sgr4], Serial_Num: last three digits, Net_Rad[W/m2]: temperature compensated net radiation, LW_Down[W/m2]: temperature compensated long wave downwelling radiation, Body_Temp[C], Ext_Volts, Scale_Factor: range and scale factor (determines number of decimal places), Operational Mode: [1 = normal], Status_Flag: [0 = normal]
Native Data Type text
Start of Message Characters
End of Message Characters
Sample Data Message 611,133,-14.18,262.67,-8.81,12.05,1,1,0
Message Parser Format ['^\d+,\d+,(?P<net_rad>\-?\d+\.\d+),(?P<lw_down>\-?\d+\.\d+),(?P<body_temp>\-?\d+\.\d+),(?P<ext_volts>\d+\.\d+),\d+,(?P<op_mode>\d+),(?P<status>\d+)$']
Data Processing
Data Processing Script NoTransformation
Data Rates
Native Sample Rate 1.0 Hz
Shipboard Network Packet Rate 1.0 Hz
Ship to Shore Rate None Hz
Data Storage & Access
Raw Data Ethernet Distribution
IP Address 10.3.0.255
Port 53136
Packet Size None
Transformed Data Ethernet Distribution
IP Address
Port None
Archive to Database True
Transmit Real-time Data to Shore True
Time Source
Source local_time_server
 

Parameters


Quantity body_temp (degrees_C) ext_volts (V) lw_down (W/m2) net_rad (W/m2) op_mode ( ) status ( )
Plot Color
 
 
 
 
 
 
Parameter Full Name Sensor Body Temperature External Voltage Longwave Downward Radiation Net Radiation op_mode Status Flag
Parameter Short Name body_temp ext_volts lw_down net_rad op_mode status
Parameter Standard Name CTMPZZ02 VOLTELEC XXXXXX XXXXXX STATFLIR STATFLIR
SAMOS Data Category Longwave Radiation
Description sensor body temperature in degrees Celsius externally supplied voltage reading temperature compensated longwave downward radiation in Watts per meter squared temperature compensated net radiation in Watts per meter squared Operational mode [1 = normal] sensor status flags [0 = normal]
Processing Status raw raw internally derived internally derived raw raw
Processing Symbol body_temp ext_volts lw_down net_rad op_mode status
Diagnostic True True False False True True
IOOS Category Meteorology Other Meteorology Meteorology Meteorology Other
Units degrees_C V W/m2 W/m2
Data Type float float float float int int
Gross Minimum Value None degrees_C None V None W/m2 None W/m2 None None
Gross Maximum Value None degrees_C None V None W/m2 None W/m2 None None
Recommended Minimum Value None degrees_C None V None W/m2 None W/m2 None None
Recommended Maximum Value None degrees_C None V None W/m2 None W/m2 None None
Global Minimum Value None degrees_C None V None W/m2 None W/m2 None None
Global Maximum Value None degrees_C None V None W/m2 None W/m2 None None
Local Minimum Value None degrees_C None V None W/m2 None W/m2 None None
Local Maximum Value None degrees_C None V None W/m2 None W/m2 None None
Lower Detection Limit None degrees_C None V None W/m2 None W/m2 None None
Saturated Value None degrees_C None V None W/m2 None W/m2 None None
Bad Data Value None degrees_C None V None W/m2 None W/m2 None None
Resolution None degrees_C None V None W/m2 None W/m2 None None
Accuracy +/- None degrees_C +/- None V +/- None W/m2 +/- None W/m2 +/- None +/- None
Precision None degrees_C None V None W/m2 None W/m2 None None
Binning Type linear linear linear linear spot spot
Online Display Format 4.1f 4.1f 4.1f 4.1f 1.0f 1.0f
Logging Modes
 

Physical Sensor Specs

Operating Limits
Input Power 5.0 - 30.0 V DC
Ideal Power None V DC
Operating Current None - 4.8 mA @ 12.0 V DC
Analog Output 0 - 1 Volts
Operating Flow Rate None - None ml s-1
Ideal Flow Rate None ml s-1
Operating Depth None - None m
Operating Temperature -40 - 80 degree_C
Lamp Life None days
Response Time 18.0 seconds
Additional Time Offset None seconds
Physical Status
Current Location installed
Current Status in use
Physical Condition excellent
Physical Configuration Mounted in CV4 Ventilation Unit
Software Version
Software Configuration
Data Acquisition Device ID
Data Acquisition Device Serial Port
Data Acquisition Device Analog/Digital Port(s)
Data Acquisition Device Modbus Unit None
Sensor Use Level primary
Enabled True
Vessel Installation Location
Vessel sikuliaq
Vessel Location Above SCR
Location Coordinates (meters)
Sensor Orientation vertical up
Height Above Waterline (meters) 9.3
Zero Line Reference (degrees from bow) None
Receives Filtered Water False
 

Maintenance & Support

Sensor Log

    Date (UTC)     Submitter     Event Type     Details        
Feb 13 2024 19:04:37     Nahorniak, Jasmine     renamed sensor     Changed sensor ID from radiom210133 to pyrgeo210133.    
Jan 30 2024 00:46:06     Nahorniak, Jasmine     sensor specs update     processing_script changed from [] to [NoTransformation]    
Jan 30 2024 00:46:06     Nahorniak, Jasmine     sensor specs update     sensor_prefix changed from [] to [pyrgeo]    
Mar 16 2023 00:05:48     Naber, Daniel     sensor specs update     current_reference_volts changed from [None] to [12.0]    
Mar 16 2023 00:05:48     Naber, Daniel     sensor specs update     current_max_milliamps changed from [None] to [4.8]    
Mar 16 2023 00:05:48     Naber, Daniel     sensor specs update     description changed from [The SGR4V has a digi ... nts or dome shading.] to [The SGR4-V has a dig ... nts or dome shading.]    
Mar 15 2023 23:13:48     Naber, Daniel     sensor specs update     response_time_seconds changed from [None] to [18.0]    
Mar 15 2023 23:13:48     Naber, Daniel     sensor specs update     temperature_max_celsius changed from [None] to [80]    
Mar 15 2023 23:13:48     Naber, Daniel     sensor specs update     temperature_min_celsius changed from [None] to [-40]    
Mar 15 2023 23:13:48     Naber, Daniel     sensor specs update     analog_units changed from [] to [Volts]    
Mar 15 2023 23:13:48     Naber, Daniel     sensor specs update     analog_max changed from [None] to [1]    
Mar 15 2023 23:13:48     Naber, Daniel     sensor specs update     analog_min changed from [None] to [0]    
Mar 15 2023 23:13:48     Naber, Daniel     sensor specs update     power_current_type changed from [AC] to [DC]    
Mar 15 2023 23:13:48     Naber, Daniel     sensor specs update     power_max_volts changed from [None] to [30.0]    
Mar 15 2023 23:13:48     Naber, Daniel     sensor specs update     power_min_volts changed from [None] to [5.0]    
Mar 15 2023 23:13:48     Naber, Daniel     sensor specs update     description changed from [The SGR4 has a digit ... nts or dome shading.] to [The SGR4V has a digi ... nts or dome shading.]    
Mar 11 2023 06:48:01     Naber, Daniel     sensor specs update     native_data_description changed from [Modbus RTU Protocol ... s_Flag: [0 = normal]] to [Modbus RTU Protocol. ... s_Flag: [0 = normal]]    
Mar 11 2023 06:43:29     Naber, Daniel     sensor specs update     maintenance_C_text changed from [test001] to []    
Mar 11 2023 06:43:29     Naber, Daniel     sensor specs update     packet_rate_hz changed from [None] to [1.0]    
Mar 11 2023 06:43:29     Naber, Daniel     sensor specs update     transmit_port changed from [None] to [53136]    
Mar 11 2023 06:43:29     Naber, Daniel     sensor specs update     transmit_ip changed from [] to [10.3.0.255]    
Mar 11 2023 06:43:29     Naber, Daniel     sensor specs update     time_source changed from [vessel_time_server] to [local_time_server]    
Mar 11 2023 06:43:29     Naber, Daniel     sensor specs update     use_level changed from [] to [primary]    
Mar 11 2023 06:43:29     Naber, Daniel     sensor specs update     height_above_waterline changed from [None] to [9.3]    
Mar 11 2023 06:43:29     Naber, Daniel     sensor specs update     sensor_orientation changed from [horizontal] to [vertical up]    
Mar 11 2023 06:43:29     Naber, Daniel     sensor specs update     text_regex_format changed from [] to [['^\d+,\d+,(?P<net_r ... ),(?P<status>\d+)$']]    
Mar 11 2023 06:43:29     Naber, Daniel     sensor specs update     sample_data_message changed from [] to [611,133,-14.18,262.67,-8.81,12.05,1,1,0]    
Mar 11 2023 06:43:29     Naber, Daniel     sensor specs update     native_type changed from [] to [text]    
Mar 11 2023 06:43:29     Naber, Daniel     sensor specs update     native_data_description changed from [Modbus RTU Protocol] to [Modbus RTU Protocol ... s_Flag: [0 = normal]]    
Mar 11 2023 06:43:29     Naber, Daniel     sensor specs update     sample_rate_hz changed from [None] to [1.0]    
Mar 11 2023 06:43:29     Naber, Daniel     sensor specs update     sensor_ethernet_protocol changed from [] to [udp]    
Mar 11 2023 06:43:29     Naber, Daniel     sensor specs update     sensor_ip changed from [] to [10.3.0.136]    
Mar 11 2023 06:43:29     Naber, Daniel     sensor specs update     serial_bytesize changed from [] to [8]    
Mar 11 2023 06:43:29     Naber, Daniel     sensor specs update     serial_stop_bits changed from [] to [1]    
Mar 11 2023 06:43:29     Naber, Daniel     sensor specs update     serial_parity changed from [] to [E]    
Mar 11 2023 06:43:29     Naber, Daniel     sensor specs update     serial_baud_rate changed from [] to [19200]    
Feb 07 2023 13:58:00     Naber, Daniel     installation     February 1, 2022    
Oct 18 2022 18:50:30     Greto, Carmen     sensor specs update     use_level changed from [primary] to []    
Oct 18 2022 18:50:30     Greto, Carmen     sensor specs update     location_text changed from [] to [Above SCR]    
Oct 18 2022 18:50:30     Greto, Carmen     sensor specs update     enabled changed from [False] to [True]    
May 23 2022 23:21:29     Greto, Carmen     sensor specs update     description changed from [Scientific pyrgeomet ... rd radiation output.] to [The SGR4 has a digit ... nts or dome shading.]    
May 23 2022 23:18:24     Greto, Carmen     sensor specs update     configuration changed from [Mounted in CV4 Ventilation Unit on SCR] to [Mounted in CV4 Ventilation Unit]    
May 23 2022 23:12:53     Greto, Carmen     new sensor     Initial submission of sensor metadata.    
 
Calibration & Maintenance

Frequency Last Completed Next Due
Maintenance A: Clear Dome of Snow and Moisture
  1 days
Maintenance B:
 
Maintenance C:
 
Calibration
  60 months Jul 19 2021 Jul 19 2026
Recommended Replacement
 
Warranty
 
 
Calibration Files

Date Version Author Type Document Access
Jul 19 2021           cal_file   Pyrgeometer SGR4_0133 CAL_PDF
 
 
User Guides & Documents

Date Version Author Type Document Access
 
 
Ownership

Quantity Value
Owner UAF
Responsible Party UAF Marine Technicians
Purchase Date Jan 01 2021
Acquire Date
Acquire Condition
End/Lost Ownership Date
 
Vendor Support

Support Type Details
Company Kipp and Zonen (OTT Hydromet)
Website https://www.kippzonen.com/
Phone +1 703 406 2800
Email kipp.usa@kippzonen.com
Shipping OTT Hydromet, 22400 Davis Dr.,Sterling, VA 20164, USA
 
Copyright Notice: All sensor-specific images and documentation on this page are copyright of the vendor.
 
 
OSU logo NSF logo RCRV logo
RCRV Datapresence and Engineering Support Center
Oregon State University
Corvallis, OR 97331, USA