PAR QSR-2150_50172

Summary



TypePAR Meter
VendorBiospherical Instruments Inc.
ModelBiospherical Instruments QSR 2150 surface reference radiometer
S/N50172
ID Prefixparqsr
CORIOLIX IDparqsr50172
ClassAtmospheric
LocationTech Stores
Conditiongood
Configuration
Native Sample Rate1.0 Hz
UDP Port(s)53104 None
EnabledFalse
 


Image Not Available
 
 

General Description



Quantum Scalar Irradiance PAR Surface Reference Sensor. Data output may be chosen between volts, volts‐offset (default), calibrated irradiance, or calibrated irradiance adjusted for immersion in water. Optionally, sensor temperature and line voltage may be included in the data stream. A preamble may be defined to the data frame to easily identify data in a multi‐sensor environment. For example "$" may be used in a NEMA application. As delivered, the output is in volts‐offset.

 
 

Data Collection Specs

Native Data Feed
Signal Type Digital
Communication Type RS-232
Serial Details
Baud Rate 19200
Parity N
Stop Bits 1
Byte Size 8
Timeout (s) None
Polled Start Commands None
Polled Serial Commands
Ethernet Details
MAC Address
IP Address 10.3.0.104
Protocol udp
Source File Details
Path
Filename
Native Data Format
Native Data Description Example Data Line: radiom50172 rad_qsr2150a 2022-02-01T22:23:11.9094Z 866.876, 12.74, 7.494 ***** Field Descriptions: Coriolix Sensor ID, LDS Logger ID, LDS Time Stamp [UTC], PAR [uE m-2 sec-1], Sensor Temperature [C], Line voltage [volts]
Native Data Type text
Start of Message Characters
End of Message Characters
Sample Data Message 0.013, 16.06, 6.137
Message Parser Format ['^(?P<par>\d+\.\d+),\s+(?P<temperature>\-?\d+\.\d+),\s+(?P<voltage>\d+\.\d+)$']
Data Processing
Data Processing Script NoTransformation
Data Rates
Native Sample Rate 1.0 Hz
Shipboard Network Packet Rate 1.0 Hz
Ship to Shore Rate None Hz
Data Storage & Access
Raw Data Ethernet Distribution
IP Address 10.3.0.255
Port 53104
Packet Size None
Transformed Data Ethernet Distribution
IP Address
Port None
Archive to Database True
Transmit Real-time Data to Shore True
Time Source
Source local_time_server
 

Parameters


Quantity par (uE/m2/s) sensor_temp (degrees_C) volts (V)
Plot Color
 
 
 
Parameter Full Name PAR Irradiance Sensor Temperature Line voltage
Parameter Short Name par sensor_temp volts
Parameter Standard Name DWIRRSSD CDTAZZ01 VOLTELEC
SAMOS Data Category Photosynthetically Active Radiation
Description PAR [uE/m2 sec] Sensor Temp [C] Sensor line voltage
Processing Status internally derived raw raw
Processing Symbol par temperature voltage
Diagnostic False True True
IOOS Category Meteorology Other Other
Units uE/m2/s degrees_C V
Data Type float float float
Gross Minimum Value None uE/m2/s None degrees_C None V
Gross Maximum Value None uE/m2/s None degrees_C None V
Recommended Minimum Value None uE/m2/s None degrees_C None V
Recommended Maximum Value None uE/m2/s None degrees_C None V
Global Minimum Value None uE/m2/s None degrees_C None V
Global Maximum Value None uE/m2/s None degrees_C None V
Local Minimum Value None uE/m2/s None degrees_C None V
Local Maximum Value None uE/m2/s None degrees_C None V
Lower Detection Limit None uE/m2/s None degrees_C None V
Saturated Value None uE/m2/s None degrees_C None V
Bad Data Value None uE/m2/s None degrees_C None V
Resolution None uE/m2/s None degrees_C None V
Accuracy +/- None uE/m2/s +/- None degrees_C +/- None V
Precision None uE/m2/s None degrees_C None V
Binning Type linear linear linear
Online Display Format 4.4f 4.2f 4.2f
Logging Modes
 

Physical Sensor Specs

Operating Limits
Input Power 6.0 - 15.0 V DC
Ideal Power None V DC
Operating Current 0.0 - 5.0 mA @ None V DC
Analog Output None - None
Operating Flow Rate None - None ml s-1
Ideal Flow Rate None ml s-1
Operating Depth 0 - None m
Operating Temperature -10 - 50 degree_C
Lamp Life None days
Response Time None seconds
Additional Time Offset None seconds
Physical Status
Current Location on vessel
Current Status not in use
Physical Condition good
Physical Configuration
Software Version
Software Configuration ADC sample rate 125, gain 1, Averaging 125 readings, Calfactor: 0.0007529 uE m-2 sec-1, Calibration Mode = B, Output calibrated for use in air dividing net volts by 0.0007529 with units of uE m-2 sec.
Data Acquisition Device ID
Data Acquisition Device Serial Port
Data Acquisition Device Analog/Digital Port(s)
Data Acquisition Device Modbus Unit None
Sensor Use Level
Enabled False
Vessel Installation Location
Vessel sikuliaq
Vessel Location Tech Stores
Location Coordinates (meters)
Sensor Orientation vertical up
Height Above Waterline (meters) 12.3
Zero Line Reference (degrees from bow) None
Receives Filtered Water False
 

Maintenance & Support

Sensor Log

    Date (UTC)     Submitter     Event Type     Details        
Sep 06 2024 18:50:07     Greto, Carmen     sensor specs update     last_cal_date changed from [2022-01-27 00:00:00+00:00] to [2024-07-31 00:00:00+00:00]    
Sep 06 2024 18:47:58     Greto, Carmen     sensor specs update     location_text changed from [] to [Tech Stores]    
Sep 06 2024 18:47:58     Greto, Carmen     sensor specs update     configuration changed from [Installed: 02-Feb-23] to []    
Sep 06 2024 18:47:58     Greto, Carmen     sensor specs update     current_status changed from [calibration in progress] to [not in use]    
Sep 06 2024 18:47:58     Greto, Carmen     sensor specs update     current_location changed from [repair/calibration] to [on vessel]    
Jun 16 2024 22:14:58     Grischuk, Jenny     sensor specs update     use_level changed from [primary] to []    
Jun 16 2024 22:14:58     Grischuk, Jenny     sensor specs update     location_text changed from [Above SCR] to []    
Jun 16 2024 22:14:58     Grischuk, Jenny     sensor specs update     enabled changed from [True] to [False]    
Jun 16 2024 22:14:58     Grischuk, Jenny     sensor specs update     current_status changed from [in use] to [calibration in progress]    
Jun 16 2024 22:14:58     Grischuk, Jenny     sensor specs update     current_location changed from [installed] to [repair/calibration]    
Feb 12 2024 20:26:59     Nahorniak, Jasmine     renamed sensor     Changed sensor ID from radqsr50172 to parqsr50172.    
Feb 12 2024 20:23:36     Nahorniak, Jasmine     renamed sensor     Changed sensor ID from radiom50172 to radqsr50172.    
Jan 29 2024 19:06:19     Nahorniak, Jasmine     sensor specs update     processing_script changed from [QSR2150] to [NoTransformation]    
Jan 29 2024 18:30:12     Nahorniak, Jasmine     sensor specs update     processing_script changed from [] to [QSR2150]    
Jan 29 2024 18:30:12     Nahorniak, Jasmine     sensor specs update     sensor_prefix changed from [] to [parqsr]    
Mar 11 2023 06:35:23     Naber, Daniel     sensor specs update     maintenance_A_freq_days changed from [None] to [1]    
Mar 11 2023 06:35:23     Naber, Daniel     sensor specs update     maintenance_A_text changed from [] to [Clear circular tray and sensor of snow]    
Mar 11 2023 06:35:23     Naber, Daniel     sensor specs update     configuration changed from [] to [Installed: 02-Feb-23]    
Mar 11 2023 06:11:05     Naber, Daniel     sensor specs update     configuration changed from [Installed: 02-Feb-23] to []    
Mar 11 2023 05:59:25     Naber, Daniel     sensor specs update     packet_rate_hz changed from [None] to [1.0]    
Mar 11 2023 05:59:25     Naber, Daniel     sensor specs update     software_configuration changed from [ADC sample rate 125, ... units of uE/m2 sec.] to [ADC sample rate 125, ... units of uE m-2 sec.]    
Mar 11 2023 05:59:24     Naber, Daniel     sensor specs update     sample_rate_hz changed from [250.0] to [1.0]    
Mar 11 2023 05:54:22     Naber, Daniel     sensor specs update     height_above_waterline changed from [None] to [12.3]    
Mar 11 2023 05:54:22     Naber, Daniel     sensor specs update     software_configuration changed from [] to [ADC sample rate 125, ... units of uE/m2 sec.]    
Mar 11 2023 05:54:22     Naber, Daniel     sensor specs update     text_regex_format changed from [] to [['^(?P<par>\d+\.\d+) ... voltage>\d+\.\d+)$']]    
Mar 11 2023 05:54:22     Naber, Daniel     sensor specs update     sample_data_message changed from [.013016, 16.06, 6.13 ... °C and line voltage] to [0.013, 16.06, 6.137]    
Mar 11 2023 05:54:22     Naber, Daniel     sensor specs update     native_data_description changed from [Data output may be c ... is in volts‐offset.] to [Example Data Line: ... Line voltage [volts]]    
Mar 11 2023 05:54:22     Naber, Daniel     sensor specs update     sensor_ethernet_protocol changed from [] to [udp]    
Mar 11 2023 05:54:22     Naber, Daniel     sensor specs update     sensor_ip changed from [] to [10.3.0.104]    
Mar 11 2023 05:54:22     Naber, Daniel     sensor specs update     serial_baud_rate changed from [9600] to [19200]    
Mar 11 2023 05:54:22     Naber, Daniel     sensor specs update     description changed from [Quantum Scalar Irrad ... ace Reference Sensor] to [Quantum Scalar Irrad ... is in volts‐offset.]    
Feb 03 2023 23:18:16     Greto, Carmen     sensor specs update     location_text changed from [Science Control Room] to [Above SCR]    
Feb 03 2023 21:24:19     Greto, Carmen     sensor specs update     transmit_port changed from [None] to [53104]    
Feb 03 2023 21:24:19     Greto, Carmen     sensor specs update     transmit_ip changed from [] to [10.3.0.255]    
Feb 03 2023 21:24:19     Greto, Carmen     sensor specs update     time_source changed from [vessel_time_server] to [local_time_server]    
Feb 03 2023 21:24:19     Greto, Carmen     sensor specs update     use_level changed from [] to [primary]    
Feb 03 2023 21:24:19     Greto, Carmen     sensor specs update     location_text changed from [Tech Stores] to [Science Control Room]    
Feb 03 2023 21:24:19     Greto, Carmen     sensor specs update     enabled changed from [False] to [True]    
Feb 03 2023 21:24:19     Greto, Carmen     sensor specs update     configuration changed from [] to [Installed: 02-Feb-23]    
Feb 03 2023 21:24:19     Greto, Carmen     sensor specs update     current_status changed from [not in use] to [in use]    
Feb 03 2023 21:24:19     Greto, Carmen     sensor specs update     current_location changed from [on vessel] to [installed]    
Dec 13 2022 00:02:00     Greto, Carmen     sensor specs update     last_cal_date changed from [2018-02-07 00:00:00+00:00] to [2022-01-27 00:00:00+00:00]    
Oct 21 2022 08:16:11     Kopka, Krystian     sensor specs update     location_text changed from [Not Applicable] to [Tech Stores]    
Oct 21 2022 00:26:50     Kopka, Krystian     sensor specs update     use_level changed from [spare] to []    
May 23 2022 23:49:33     Greto, Carmen     sensor specs update     sensor_name changed from [PAR_QSR-2150_50172] to [PAR QSR-2150_50172]    
Dec 16 2021 21:08:48     Greto, Carmen     sensor specs update     current_max_milliamps changed from [None] to [5.0]    
Dec 16 2021 21:08:48     Greto, Carmen     sensor specs update     current_min_milliamps changed from [None] to [0.0]    
Dec 13 2021 20:08:35     Greto, Carmen     sensor specs update     time_source changed from [] to [vessel_time_server]    
Dec 13 2021 20:08:35     Greto, Carmen     sensor specs update     use_level changed from [] to [spare]    
Dec 13 2021 20:08:35     Greto, Carmen     sensor specs update     sensor_orientation changed from [] to [vertical up]    
Dec 13 2021 20:08:35     Greto, Carmen     sensor specs update     location_text changed from [] to [Not Applicable]    
Dec 13 2021 20:08:35     Greto, Carmen     sensor specs update     condition changed from [] to [good]    
Dec 13 2021 20:08:35     Greto, Carmen     sensor specs update     current_status changed from [] to [not in use]    
Dec 13 2021 20:08:35     Greto, Carmen     sensor specs update     current_location changed from [] to [on vessel]    
Dec 13 2021 20:08:35     Greto, Carmen     sensor specs update     temperature_max_celsius changed from [None] to [50]    
Dec 13 2021 20:08:35     Greto, Carmen     sensor specs update     temperature_min_celsius changed from [None] to [-10]    
Dec 13 2021 20:08:35     Greto, Carmen     sensor specs update     depth_min_meters changed from [None] to [0]    
Dec 13 2021 20:08:35     Greto, Carmen     sensor specs update     power_current_type changed from [] to [DC]    
Dec 13 2021 20:08:35     Greto, Carmen     sensor specs update     power_max_volts changed from [None] to [15]    
Dec 13 2021 20:08:35     Greto, Carmen     sensor specs update     power_min_volts changed from [None] to [6]    
Dec 13 2021 20:08:35     Greto, Carmen     sensor specs update     sample_data_message changed from [] to [.013016, 16.06, 6.13 ... °C and line voltage]    
Dec 13 2021 20:08:35     Greto, Carmen     sensor specs update     native_type changed from [] to [text]    
Dec 13 2021 20:08:35     Greto, Carmen     sensor specs update     native_data_description changed from [] to [Data output may be c ... is in volts‐offset.]    
Dec 13 2021 20:08:35     Greto, Carmen     sensor specs update     sample_rate_hz changed from [None] to [250.0]    
Dec 13 2021 20:08:35     Greto, Carmen     sensor specs update     serial_bytesize changed from [] to [8]    
Dec 13 2021 20:08:35     Greto, Carmen     sensor specs update     serial_stop_bits changed from [] to [1]    
Dec 13 2021 20:08:35     Greto, Carmen     sensor specs update     serial_parity changed from [] to [N]    
Dec 13 2021 20:08:35     Greto, Carmen     sensor specs update     serial_baud_rate changed from [] to [9600]    
Dec 13 2021 20:08:35     Greto, Carmen     sensor specs update     comm_type changed from [] to [RS-232]    
Dec 13 2021 20:08:35     Greto, Carmen     sensor specs update     signal_type changed from [] to [Digital]    
Dec 13 2021 20:08:35     Greto, Carmen     sensor specs update     description changed from [Reference PAR] to [Quantum Scalar Irrad ... ace Reference Sensor]    
Dec 13 2021 20:08:35     Greto, Carmen     sensor specs update     model changed from [QSR-2150] to [Biospherical Instrum ... reference radiometer]    
Dec 13 2021 20:08:35     Greto, Carmen     sensor specs update     vendor changed from [Biospherical Instruments Inc.] to [Biospherical Instruments Inc. ]    
Dec 13 2021 20:08:35     Greto, Carmen     sensor specs update     sensor_name changed from [Reference PAR] to [PAR_QSR-2150_50172]    
Feb 17 2021 02:24:58     Roberts, Steve     sensor specs update     vessel changed from [] to [Sikuliaq]    
Feb 17 2021 02:23:24     Roberts, Steve     new sensor     Initial submission of sensor metadata.    
 
Calibration & Maintenance

Frequency Last Completed Next Due
Maintenance A: Clear circular tray and sensor of snow
  1 days
Maintenance B:
 
Maintenance C:
 
Calibration
  24 months Jul 31 2024 Jul 31 2026
Recommended Replacement
 
Warranty
 
 
Calibration Files

Date Version Author Type Document Access
Jul 31 2024        CJG   cal_file   PAR QSR-2150_50172 CAL_PDF
Jan 27 2022           cal_file   PAR QSR-2150_50172 CAL_PDF 27JAN22
Feb 07 2018           cal_file   PAR QSR-2150_50172 CAL_PDF
Apr 30 2013           cal_file   PAR QSR-2150_50172 CAL_PDF
 
 
User Guides & Documents

Date Version Author Type Document Access
 
 
Ownership

Quantity Value
Owner UAF
Responsible Party UAF Marine Technicians
Purchase Date
Acquire Date
Acquire Condition
End/Lost Ownership Date
 
Vendor Support

Support Type Details
Company Biospherical Instruments Inc.
Website http://www.biospherical.com
Phone +1 619 686 1888
Email support@biospherical.com
Shipping Biospherical Instruments Inc., 5340 Riley St., San Diego, CA 92110, USA
 
Copyright Notice: All sensor-specific images and documentation on this page are copyright of the vendor.
 
 
OSU logo NSF logo RCRV logo
RCRV Datapresence and Engineering Support Center
Oregon State University
Corvallis, OR 97331, USA