Oxygen Aanderaa_3594

Summary



TypeOxygen Sensor
VendorAanderaa
ModelAanderaa Oxygen Optode 4330 and 4330F
S/N3594
ID Prefixoxygen
CORIOLIX IDoxygen003594
ClassFlowthrough
LocationMain Lab / Wet Wall
Conditionexcellent
ConfigurationWater source: bow intake UNCSW system (6.0 meters below average waterline) or centerboard UNCSW system if bow intake is clogged with ice. Refer to wh300_xducer_depth or R2R Event Logger for position/depth of the centerboard.
Native Sample Rate0.2 Hz
UDP Port(s)54132 None
EnabledTrue
 


Image Not Available
 
 

General Description



The Oxygen Optode is an optical sensor that does not consume oxygen. The measurement principle is based on fluorescence quenching, while traditional polorographic oxygen sensors, often called Clark sensors, based on electrochemical principles consume oxygen.

 
 

Data Collection Specs

Native Data Feed
Signal Type Digital
Communication Type RS-232
Serial Details
Baud Rate 9600
Parity N
Stop Bits 1
Byte Size 8
Timeout (s) None
Polled Serial Commands
Ethernet Details
MAC Address
IP Address 10.3.0.132
Protocol udp
Source File Details
Path
Filename
Native Data Format
Native Data Description The oxygen optode outputs data in AiCaP CANbus and RS-232. The sensor can present the O2 concentration in μM, the air saturation in % and the temperature in °C. Includes salinity compensated O2 Concentration using TSG-sbe45 temperature and salinity. Example data line: oxygen_optode4330_cor 2021-05-24T00:00:03.1008Z 4330,3594,503.7600,408.7601,109.8400,-0.1550,0.0000,-0.1741,30.1824,62.012229,-168.620039 Field Descriptions: Log Name Log Time ModelNo. SerialNo. O2Concentration[uM] salinty corrected O2Concentration[uM] AirSaturation[%] Optode Temp[C] reference salinity[psu] TSG Temp[C] TSG Salinity[psu] latitude longitude
Native Data Type text
Start of Message Characters
End of Message Characters
Sample Data Message 4330,3594,503.7600,408.7601,109.8400,-0.1550,0.0000,-0.1741,30.1824,62.012229,-168.620039
Message Parser Format ['^\d+,\d+,(?P<oxy_concentration>\d+\.\d+),(?P<oxy_concentration_corr>\d+\.\d+),(?P<air_saturation>\d+\.\d+),(?P<temp>\-?\d+\.\d+),(?P<salinity_ref>\d+\.\d+),(?P<tsg_temp>\-?\d+\.\d+),(?P<tsg_sal>\d+\.\d+),(?P<latitude>\-?\d+\.\d+),(?P<longitude>\-?\d+\.\d+)$']
Data Processing
Data Processing Script NoTransformation
Data Rates
Native Sample Rate 0.2 Hz
Shipboard Network Packet Rate 0.2 Hz
Ship to Shore Rate None Hz
Data Storage & Access
Raw Data Ethernet Distribution
IP Address 10.3.0.255
Port 54132
Packet Size None
Transformed Data Ethernet Distribution
IP Address
Port None
Archive to Database True
Transmit Real-time Data to Shore True
Time Source
Source local_time_server
 

Parameters


Quantity Air Saturation (%) Oxygen (umol/l) Oxygen Corr (umol/l) Temperature (degrees_C)
Plot Color
 
 
 
 
Parameter Full Name Air Saturation Oxygen Concentration Oxygen Concentration Corrected Temperature
Parameter Short Name Air Saturation Oxygen Oxygen Corr Temperature
Parameter Standard Name XXXXXX XXXXXX XXXXXX XXXXXX
SAMOS Data Category
Description
Processing Status raw internally derived internally derived raw
Processing Symbol air_saturation oxy_concentration oxy_concentration_corr temp
Diagnostic False False False True
IOOS Category Dissolved O2 Dissolved O2 Dissolved O2 Other
Units % umol/l umol/l degrees_C
Data Type float float float float
Gross Minimum Value None % None umol/l None umol/l None degrees_C
Gross Maximum Value None % None umol/l None umol/l None degrees_C
Recommended Minimum Value None % None umol/l None umol/l None degrees_C
Recommended Maximum Value None % None umol/l None umol/l None degrees_C
Global Minimum Value None % None umol/l None umol/l None degrees_C
Global Maximum Value None % None umol/l None umol/l None degrees_C
Local Minimum Value None % None umol/l None umol/l None degrees_C
Local Maximum Value None % None umol/l None umol/l None degrees_C
Lower Detection Limit None % None umol/l None umol/l None degrees_C
Saturated Value None % None umol/l None umol/l None degrees_C
Bad Data Value None % None umol/l None umol/l None degrees_C
Resolution None % None umol/l None umol/l None degrees_C
Accuracy +/- None % +/- None umol/l +/- None umol/l +/- None degrees_C
Precision None % None umol/l None umol/l None degrees_C
Binning Type linear linear linear linear
Online Display Format 4.1f 4.1f 4.1f 4.1f
Logging Modes
 

Physical Sensor Specs

Operating Limits
Input Power 5.0 - 14.0 V DC
Ideal Power None V DC
Operating Current 0.16 - 100.0 mA @ 12.0 V DC
Analog Output None - None
Operating Flow Rate None - None ml s-1
Ideal Flow Rate None ml s-1
Operating Depth 0 - 300 m
Operating Temperature -5 - 40 degree_C
Lamp Life None days
Response Time 2.0 seconds
Additional Time Offset None seconds
Physical Status
Current Location installed
Current Status in use
Physical Condition excellent
Physical Configuration Water source: bow intake UNCSW system (6.0 meters below average waterline) or centerboard UNCSW system if bow intake is clogged with ice. Refer to wh300_xducer_depth or R2R Event Logger for position/depth of the centerboard.
Software Version
Software Configuration
Data Acquisition Device ID
Data Acquisition Device Serial Port
Data Acquisition Device Analog/Digital Port(s)
Data Acquisition Device Modbus Unit None
Sensor Use Level primary
Enabled True
Vessel Installation Location
Vessel Sikuliaq
Vessel Location Main Lab / Wet Wall
Location Coordinates (meters)
Sensor Orientation horizontal fwd
Height Above Waterline (meters) None
Zero Line Reference (degrees from bow) None
Receives Filtered Water False
 

Maintenance & Support

Sensor Log

    Date (UTC)     Submitter     Event Type     Details        
Apr 24 2024 15:30:33     Grischuk, Jenny     sensor specs update     maintenance_B_freq_days changed from [None] to [183]    
Apr 24 2024 15:30:33     Grischuk, Jenny     sensor specs update     maintenance_B_text changed from [] to [Cleaning the optode housing]    
Apr 24 2024 15:30:00     Grischuk, Jenny     maintenance B     cleaned housing, very little sediment found    
Feb 16 2024 20:29:51     Nahorniak, Jasmine     new parameter     Copied Oxygen (umol/l) parameter metadata from oxygen003594.    
Feb 16 2024 20:28:22     Nahorniak, Jasmine     sensor specs update     text_regex_format changed from [['^\d+\s+\d+\s+(?P<o ... aw_temp>\d+\.\d+)$']] to [['^\d+,\d+,(?P<oxy_c ... tude>\-?\d+\.\d+)$']]    
Feb 16 2024 20:28:22     Nahorniak, Jasmine     sensor specs update     sample_data_message changed from [4330 3594 222. ... 68.1 843.7 155.3] to [4330,3594,503.7600,4 ... 2.012229,-168.620039]    
Feb 16 2024 20:28:22     Nahorniak, Jasmine     sensor specs update     native_data_description changed from [The oxygen optode ou ... p[mV], RawTemp[mV]] to [The oxygen optode ou ... atitude longitude]    
Feb 16 2024 03:37:03     Nahorniak, Jasmine     sensor specs update     processing_script changed from [] to [NoTransformation]    
Feb 16 2024 03:37:03     Nahorniak, Jasmine     sensor specs update     sensor_prefix changed from [] to [oxygen]    
Dec 16 2023 23:23:04     Naber, Daniel     sensor specs update     configuration changed from [Water source: bow in ... epth of centerboard.] to [Water source: bow in ... of the centerboard.]    
Dec 16 2023 01:24:32     Naber, Daniel     sensor specs update     configuration changed from [Water source: bow in ... epth of centerboard.] to [Water source: bow in ... epth of centerboard.]    
Mar 22 2023 05:07:58     Naber, Daniel     sensor specs update     configuration changed from [Water source: bow in ... epth of centerboard.] to [Water source: bow in ... epth of centerboard.]    
Mar 10 2023 06:59:31     Naber, Daniel     sensor specs update     maintenance_C_text changed from [test001] to []    
Mar 10 2023 06:59:31     Naber, Daniel     sensor specs update     maintenance_A_text changed from [] to [Fresh water flush]    
Mar 10 2023 06:59:31     Naber, Daniel     sensor specs update     transmit_rate_hz changed from [1.0] to [None]    
Mar 10 2023 06:59:31     Naber, Daniel     sensor specs update     packet_rate_hz changed from [1.0] to [0.2]    
Mar 10 2023 06:59:31     Naber, Daniel     sensor specs update     location_text changed from [Main Lab] to [Main Lab / Wet Wall]    
Mar 10 2023 06:59:31     Naber, Daniel     sensor specs update     configuration changed from [Plumbed after the Pr ... ain of the Wet Wall.] to [Water source: bow in ... epth of centerboard.]    
Mar 10 2023 06:59:31     Naber, Daniel     sensor specs update     text_regex_format changed from [] to [['^(?<model_number>\ ... aw_temp>\d+\.\d+)$']]    
Mar 10 2023 06:59:31     Naber, Daniel     sensor specs update     sample_data_message changed from [4330 3594 222. ... Amp[mV] RawTemp[mV]] to [4330 3594 222. ... 68.1 843.7 155.3]    
Mar 10 2023 06:59:31     Naber, Daniel     sensor specs update     native_data_description changed from [The oxygen optode ou ... e temperature in °C.] to [The oxygen optode ou ... p[mV], RawTemp[mV]]    
Feb 07 2023 13:44:00     Naber, Daniel     installation     February 21, 2021    
Feb 06 2023 04:02:03     Naber, Daniel     sensor specs update     transmit_port changed from [None] to [54132]    
Feb 06 2023 04:02:03     Naber, Daniel     sensor specs update     transmit_ip changed from [] to [10.3.0.255]    
Feb 06 2023 04:02:03     Naber, Daniel     sensor specs update     time_source changed from [vessel_time_server] to [local_time_server]    
Feb 06 2023 04:02:03     Naber, Daniel     sensor specs update     use_level changed from [] to [primary]    
Feb 06 2023 04:02:03     Naber, Daniel     sensor specs update     sensor_orientation changed from [horizontal aft] to [horizontal fwd]    
Feb 06 2023 04:02:03     Naber, Daniel     sensor specs update     sensor_ethernet_protocol changed from [] to [udp]    
Feb 06 2023 04:02:03     Naber, Daniel     sensor specs update     sensor_ip changed from [] to [10.3.0.132]    
Oct 21 2022 00:11:13     Kopka, Krystian     sensor specs update     use_level changed from [primary] to []    
Oct 18 2022 18:48:08     Greto, Carmen     sensor specs update     enabled changed from [False] to [True]    
Oct 07 2022 09:28:27     Kopka, Krystian     sensor specs update     last_cal_date changed from [2020-09-12 00:00:00+00:00] to [2021-01-08 00:00:00+00:00]    
Sep 07 2022 18:39:08     Greto, Carmen     sensor specs update     sensor_name changed from [O2 Aanderaa_3594] to [Oxygen Aanderaa_3594]    
May 23 2022 23:47:47     Greto, Carmen     sensor specs update     sensor_name changed from [O2_Aanderaa_3594] to [O2 Aanderaa_3594]    
May 23 2022 23:24:21     Greto, Carmen     sensor specs update     enabled changed from [True] to [False]    
Jan 28 2022 18:10:22     Greto, Carmen     sensor specs update     sensor_name changed from [O2_Optode_3594] to [O2_Aanderaa_3594]    
Jan 28 2022 18:09:52     Greto, Carmen     sensor specs update     sensor_name changed from [Oxygen Optode] to [O2_Optode_3594]    
Jan 28 2022 18:09:01     Greto, Carmen     sensor specs update     last_cal_date changed from [None] to [2020-09-12 00:00:00+00:00]    
Jan 28 2022 18:09:01     Greto, Carmen     sensor specs update     transmit_rate_hz changed from [None] to [1.0]    
Jan 28 2022 18:09:01     Greto, Carmen     sensor specs update     packet_rate_hz changed from [None] to [1.0]    
Jan 28 2022 18:09:01     Greto, Carmen     sensor specs update     time_source changed from [] to [vessel_time_server]    
Jan 28 2022 18:09:01     Greto, Carmen     sensor specs update     use_level changed from [] to [primary]    
Jan 28 2022 18:09:01     Greto, Carmen     sensor specs update     location_text changed from [Flowthrough] to [Main Lab]    
Jan 28 2022 18:09:01     Greto, Carmen     sensor specs update     acquire_condition changed from [] to [excellent]    
Jan 28 2022 18:09:01     Greto, Carmen     sensor specs update     purchase_date changed from [None] to [2021-01-01 00:00:00+00:00]    
Jan 28 2022 18:09:00     Greto, Carmen     sensor specs update     responsible_party changed from [other] to [UAF Marine Technicians]    
Jan 28 2022 18:09:00     Greto, Carmen     sensor specs update     owner changed from [other] to [UAF]    
Jan 28 2022 18:09:00     Greto, Carmen     sensor specs update     response_time_seconds changed from [None] to [2.0]    
Jan 28 2022 18:09:00     Greto, Carmen     sensor specs update     temperature_max_celsius changed from [None] to [40]    
Jan 28 2022 18:09:00     Greto, Carmen     sensor specs update     temperature_min_celsius changed from [None] to [-5]    
Jan 28 2022 18:09:00     Greto, Carmen     sensor specs update     depth_max_meters changed from [None] to [300]    
Jan 28 2022 18:09:00     Greto, Carmen     sensor specs update     depth_min_meters changed from [None] to [0]    
Jan 28 2022 18:09:00     Greto, Carmen     sensor specs update     current_reference_volts changed from [None] to [12.0]    
Jan 28 2022 18:09:00     Greto, Carmen     sensor specs update     current_max_milliamps changed from [None] to [100.0]    
Jan 28 2022 18:09:00     Greto, Carmen     sensor specs update     current_min_milliamps changed from [None] to [0.16]    
Jan 28 2022 18:09:00     Greto, Carmen     sensor specs update     power_current_type changed from [] to [DC]    
Jan 28 2022 18:09:00     Greto, Carmen     sensor specs update     sample_data_message changed from [] to [4330 3594 222. ... Amp[mV] RawTemp[mV]]    
Jan 28 2022 18:09:00     Greto, Carmen     sensor specs update     native_type changed from [] to [text]    
Jan 28 2022 18:09:00     Greto, Carmen     sensor specs update     native_data_description changed from [] to [The oxygen optode ou ... e temperature in °C.]    
Jan 28 2022 18:09:00     Greto, Carmen     sensor specs update     sample_rate_hz changed from [None] to [0.2]    
Jan 28 2022 18:09:00     Greto, Carmen     sensor specs update     model changed from [4330] to [Aanderaa Oxygen Optode 4330 and 4330F]    
Jan 28 2022 18:09:00     Greto, Carmen     sensor specs update     vendor changed from [Aanderaa] to [Aanderaa ]    
Feb 20 2021 21:25:48     Naber, Daniel     new sensor     Initial submission of sensor metadata.    
 
Calibration & Maintenance

Frequency Last Completed Next Due
Maintenance A: Fresh water flush
 
Maintenance B: Cleaning the optode housing
  183 days Apr 24 2024
Maintenance C: test001
 
Calibration
  Jan 08 2021
Recommended Replacement
 
Warranty
 
 
Calibration Files

Date Version Author Type Document Access
Sep 12 2020           cal_file   Oxygen Aanderaa_3594 CAL_PDF
 
 
User Guides & Documents

Date Version Author Type Document Access
Jan 01 2020        Aanderaa   user_manual   O2 Aanderaa Optode Manual
Jan 01 2020        Aanderaa   datasheet   O2 Aanderaa Optode Model 4330 Datasheet
Jan 08 2021           datasheet   Oxygen Aanderaa_s/n-3594_DataSheet
Jan 08 2021           reference_sheet   Oxygen Aanderaa_s/n-3594_Pres_Test
 
 
Ownership

Quantity Value
Owner UAF
Responsible Party UAF Marine Technicians
Purchase Date Jan 01 2021
Acquire Date
Acquire Condition excellent
End/Lost Ownership Date
 
Vendor Support

Support Type Details
Company Aanderaa
Website https://www.aanderaa.com/
Phone 1 727-565-2201
Email aanderaa.us-sales@xyleminc.com
Shipping YSI Attn: Jon Fajans 9843 18th St North, Suite 1200, St. Petersburg, FL 33716 USA
 
Copyright Notice: All sensor-specific images and documentation on this page are copyright of the vendor.
 
 
Data Visualization

Time-Series Plots
   Flowthrough
   Meteorological
   Navigation
   Combination
   Custom
Charting Tools
 
Data Access

Data Structure
Quality Flags
Long-term Archive (R2R)
Data Download
   Full Resolution
   Binned Data
Web Services
   Direct Access Overview
   RESTful API
   ERDDAP
Software Tools
   Python Tools
   Matlab Tools
   R Tools
   ArcGIS Tools
 
Science Party Tools

Cruise & Participants Lists
Event Log
 
OSU logo NSF logo RCRV logo